Science Inventory

Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method

Citation:

Nelson, G., J. Guynn, AND B. Chorley. Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method. Journal of Visualized Experiments . JoVE, Somerville, MA, 127:e55911, (2017). https://doi.org/10.3791/55911

Impact/Purpose:

The article is a methods paper based on a general protocol for a capillary-based immunoassay that was a key tool for a recent publication from our lab (Currier et al. PLOS ONE 2016; PMID: 27195669). While the research presented in this manuscript was optimization data performed as a result of this paper, which supported ACE PEP research, the method itself is a general tool and potential replacement technology for standard Western blot immunoassays of proteins in prepared samples. Therefore the method has wide applicability to RAP-related research. Of note, the target journal for this paper is the Journal of Visualized Experiments (JoVE). If accepted, the protocol will be published as a visual protocol and highlight the EPA authors (Chorley, Guynn, and Nelson). This is therefore not only a valuable scientific reference, but highlights the technology utilized by EPA scientists to support our Agency work.

Description:

New technologies that utilize capillary-based immunoassays promise faster and more quantitative protein assessment compared to traditional western blot procedures. However, similar to western blot analysis, optimization of capillary-based immunoassay parameters, such as protein concentration, antibody dilution, and exposure time is an important prerequisite to the generation of meaningful and reliable data. Quantitative measurements must fall within the linear range of the assay where changes in signal are directly proportional to changes in lysate concentration. Here we demonstrate the process of choosing appropriate lysate concentrations, antibody dilutions, and exposure times in the human bronchial epithelial cell line, BEAS-2B. Assay linearity is shown over a range of lysate concentrations with p53 and ɑ-tubulin antibodies. An example of signal burnout is seen at the highest concentrations with long exposure times, and an α-tubulin antibody dilution curve is shown demonstrating saturation. In addition, example experimental results are reported for doxorubicin-treated cells using our optimized parameters.

Record Details:

Record Type:DOCUMENT( JOURNAL/ PEER REVIEWED JOURNAL)
Product Published Date:09/10/2017
Record Last Revised:10/23/2018
OMB Category:Other