Science Inventory

Laboratory and field based evaluation of chromatography related performance of the Monitor for AeRosols and Gases in ambient Air (MARGA)

Citation:

Chen, X. AND Johnt Walker. Laboratory and field based evaluation of chromatography related performance of the Monitor for AeRosols and Gases in ambient Air (MARGA). 2017 AAAR, Raleigh, North Carolina, October 16 - 20, 2017.

Impact/Purpose:

MARGA’s capability of near real-time (hourly) simultaneous measurement of water soluble particulate species as well as their gaseous precursors makes it a state-of-art research instrument. Such time-resolved measurements allow investigation of highly time sensitive, rapidly changing pollution episodes as well as aerosol processes such as gas/particle partitioning and neutralization state. The MARGA has been deployed in widely varying environments to monitor ambient gaseous and particulate water soluble species including NH3, SO2, HNO3, NH4+, SO42- and NO3-. Though differences between the MARGA and other measurement systems have been observed, the extent to which the differences may be attributable solely to chromatography has not been evaluated.

Description:

The semi-continuous Monitor for AeRosols and Gases in Ambient air (MARGA) was evaluated using laboratory and field data with a focus on chromatography. The performance and accuracy assessment revealed various errors and uncertainties resulting from mis-identification and mis-integration of chromatogram peaks by MARGA automated software. To aid data reprocessing efficiency and flexibility, an alternative chromatography data processing software was adopted to further evaluate MARGA generated data for method detection limits as well as accuracy and precision. Such reprocessing and calibration significantly improved data quality by lowering method detection limits (by a factor of 1.5) and reducing variability between parallel sampler boxes. Further evaluation of instrument performance, including examination of diurnal patterns of observed gaseous and particulate water soluble species (NH3, SO2, HNO3, NH4+, SO42- and NO3-) as well fine particle neutralization state, was conducted during an intensive field campaign. The evaluation of MARGA chromatography software revealed various potential issues encountered during monitoring especially when ambient concentrations were low. Under such circumstances a bias as high as 30% could be associated with un-examined and un-screened MARGA datasets in addition to invalid data due to peak mis-identification and mis-integration. Furthermore, calibration and verification of accuracy by external standards is recommended for additional quality control given the current reliance on a single point internal standard. Analysis of field measurements across different synoptic meteorological patterns, including a NO3- rich arctic air mass, demonstrate that the MARGA is capable of resolving rapid changes in PM2.5 composition.

URLs/Downloads:

AAAR 2017 CHEN AND WALKER_MARGA (1).PDF  (PDF, NA pp,  35.734  KB,  about PDF)

Record Details:

Record Type:DOCUMENT( PRESENTATION/ ABSTRACT)
Product Published Date:10/20/2017
Record Last Revised:10/24/2017
OMB Category:Other
Record ID: 338005