Science Inventory

DISSOLVED OXYGEN AND METHANE IN WATER BY A GC HEADSPACE EQUILIBRATION TECHNIQUE

Citation:

Kampbell*, D H. AND J T. Wilson*. DISSOLVED OXYGEN AND METHANE IN WATER BY A GC HEADSPACE EQUILIBRATION TECHNIQUE. INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY 36(4):249-257, (1989).

Impact/Purpose:

Information.

Description:

An analytical procedure is described for the determination of dissolved oxygen and methane in groundwater samples. The method consists of generating a helium gas headspace in a water filled bottle, and analysis of the headspace by gas chromatography. Other permanent gases such as nitrogen, and volatile aliphatic hydrocarbons such as ethane, propane, and butane could also be analyzed. BTX analyses could also be done on the sample. Detection limit for oxygen was 0.1 mg/l and 0.002 mg/l for methane. Good agreement was shown between Winkler titration and the GC-Headspace Equilibration Technique for oxygen analyses by a linear regression coefficient, R 2 = 0.998. Oxygen was greatly depleted in some field samples when they were stored for 30 days at 4°C without hydrochloric acid preservation.

Record Details:

Record Type:DOCUMENT( JOURNAL/ PEER REVIEWED JOURNAL)
Product Published Date:08/01/1989
Record Last Revised:03/20/2009
OMB Category:Other
Record ID: 128716