Experimental assessment of elemental analyzer isotope ratio mass spectrometry normalization methodologies for environmental stable isotopes

Citation:

Balint, S., M. Schwartz, D. Fowler, S. Linnekogel, S. Cromratie Clemons, AND L. Burkemper. Experimental assessment of elemental analyzer isotope ratio mass spectrometry normalization methodologies for environmental stable isotopes. Rapid Communications in Mass Spectrometry. Wiley InterScience, Silver Spring, MD, 38(17):e9837, (2024). https://doi.org/10.1002/rcm.9837

Impact/Purpose:

This work explores and describes common methods for normalizing sample data to known standard values after analysis on an isotope ratio mass spectrometer. While may researchers and laboratories use this instrument, and the data that result from it, there are no common normalization techniques, nor have best practices been established. This study corrects this oversight and provides analysts with guidance for how to normalize their standard unknowns to standards and thus increase the accuracy and precision of their results.

Description:

Rationale In stable isotope mass spectrometry, isotope delta values are normalized to internationally recognized reference scales using a combination of certified and in-house isotope reference materials. Numerous techniques exist for performing this normalization, but these methodologies need to be experimentally assessed to compare their effect on reproducibility of isotope results. Methods We tested normalization methods by the number of reference materials used, their matrix, their isotope range, and whether normalization required extrapolating beyond the isotope range. We analyzed eight commercially available isotope reference materials on a ThermoFinnigan Delta-V isotope ratio mass spectrometer (IRMS) and an Elementar VisION IRMS for nitrogen and carbon isotope composition via solid combustion with an elemental analyzer and computed every possible isotope normalization (n = 612). Additionally, we assessed how sample matrix affected linearity effects on both instruments using five in-house reference materials. Results Normalizations exhibited the best performance when the reference materials spanning an isotope range of at least 20‰ were matrix matched with the samples and did not require extrapolation beyond the calibration curve. When these conditions were not met, the number of reference materials used had a significant effect on accuracy, with normalizations composed of two reference materials exhibiting particularly inconsistent performance at isotope ranges below 20‰. Linearity effects were found to exceed instrument precision by two orders of magnitude irrespective of matrix type and were not predicted by working gas diagnostics. Conclusions Interlaboratory comparability of isotope results is improved when operators of elemental analyzer isotope ratio mass spectrometry (EAIRMS) systems select reference materials spanning an isotope range of at least 20‰. Additionally, using three or more isotopic reference materials, avoiding extrapolation beyond the range of the normalization curve, and matching the matrix of the reference materials to the samples improve normalizations.

Record Details:

Record Type:DOCUMENT( JOURNAL/ PEER REVIEWED JOURNAL)
Product Published Date:09/15/2024
Record Last Revised:06/26/2024
OMB Category:Other
Record ID: 361923