Full Record Display for the EPA National Library Catalog

RECORD NUMBER: 2 OF 16

Main Title Modification of methods 9030 and 9031 for the analysis of sulfide by specific ion electrode /
Author Hillman, Daniel C.
Other Authors
Author Title of a Work
Nowinksi, Piotr.
Pyle, Steven M.
CORP Author Lockheed Engineering and Sciences Co., Inc., Las Vegas, NV.;Environmental Monitoring Systems Lab., Las Vegas, NV.
Publisher U.S. Environmental Protection Agency, Office of Research and Development, Environmental Monitoring Systems Laboratory ; [Distributed by the National Technical Information Service],
Year Published 1990
Report Number EPA 600-4-90-024; 68-03-3249; EPA-68-03-3249
Stock Number PB90274135
OCLC Number 37395851
Subjects Sulfides--Analysis ; Soils--Analysis ; Sulphides--Analysis
Additional Subjects Hydrogen sulfide ; Chemical analysis ; Precision ; Accuracy ; Experimental design ; Chemical reactions ; Electrochemistry ; Precipitation(Chemistry) ; Scrubbers ; Ion selective electrode analysis ; Liquid wastes ; Sulfides ; Solid wastes ; EPA method 9031 ; EPA method 9030 ; Soil contamination ; Laboratory tests
Internet Access
Description Access URL
https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=P100VKDJ.PDF
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
EJBD  EPA 600-4-90-024 Headquarters Library/Washington,DC 05/19/2020
ELBD ARCHIVE EPA 600-4-90-024 Received from HQ AWBERC Library/Cincinnati,OH 10/04/2023
NTIS  PB90-274135 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation viii, 51 pages : illustrations ; 28 cm
Abstract Two OSW SW-846 methods used for the determination of sulfide have been modified to include the use of sulfide specific ion electrodes (SIE). Currently in both methods sulfide is converted to hydrogen sulfide and distilled into a scrubber solution for subsequent determination by iodometric titration. In the modified methods, the hydrogen sulfide in the scrubber is determined by sulfide SIE. A single lab evaluation was performed to determine the operating characteristics. The sulfide SIE is linear over the range 0.25-6000 mg/L sulfide with a detection limit is about 0.2 mg/L sulfide. Over the range 5-6000 mg/L, the relative precision of the SIE is 2-4 percent. The accuracy (expressed as percent recovery) over the range 0.25-6000 mg/L varies from 75-103 percent. The sulfide SIE is very selective for the sulfide dianion and in the scrubber solution, there are no interferences. Recoveries in real samples spiked with 17.5 mg/L sulfide varied from 68-77 percent before distillation and 93-98 percent after distillation. The results from the evaluation indicate that the sulfide SIE provides an alternate technique to determine sulfide in environmental samples after distillation.
Notes "Project officer, Steven M. Pyle." "Contract Number 68-03-3249." "September 1990"--Cover Includes bibliographical references. "EPA 600-4-90-024." Cover title.
Place Published Las Vegas, Nev. Springfield, Va.
Supplementary Notes Sponsored by Environmental Monitoring Systems Lab., Las Vegas, NV.
Corporate Au Added Ent Environmental Monitoring Systems Laboratory (Las Vegas, Nev.); Lockheed Engineering & Sciences Company.
PUB Date Free Form 1990
NTIS Prices PC A04/MF A01
BIB Level m
Medium unmediated
Content text
Carrier volume
Cataloging Source OCLC/T
OCLC Time Stamp 20190726054147
Language eng
Origin OCLC
Type MERGE
OCLC Rec Leader 01933cam 2200457Ia 45010