Full Record Display for the EPA National Library Catalog
RECORD NUMBER: 2 OF 3Main Title | Evaluation of the Toshiba Capacitor Discharge Mobile X-Ray Machine. | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Author | Grove, J, G. F. ; | |||||||||||
CORP Author | Bureau of Radiological Health, Rockville, Md. Div. of Electronic Products. | |||||||||||
Year Published | 1971 | |||||||||||
Report Number | BRH/DEP-72-1; 72-8006; | |||||||||||
Stock Number | PB-202 925 | |||||||||||
Additional Subjects | ( X-ray apparatus ; Evaluation) ; ( Radiation hazards ; X-ray apparatus) ; Safety engineering ; Mobile equipment ; Spurious radiation ; Capacitor discharge X-ray machines | |||||||||||
Holdings |
|
|||||||||||
Collation | 30p | |||||||||||
Abstract | Tests were performed on a Toshiba KCD-10M-6 capacitor discharge mobile x-ray machine to determine accuracy of voltage indications, the relationship between voltage setting and current-time values, line voltage effects and current drain, exposure reproducibility, focal spot size and location, correspondence between the collimator and light localizer, inherent filtration, half-value layer, radiation output, and tube housing and collimator leakage. Spurious radiation was found to originate in the tube under certain conditions. These conditions were investigated, and the observed radiation levels are reported. (Author) | |||||||||||
Supplementary Notes | Paper copy available from GPO $0.35 as stock no. 1715-0024. | |||||||||||
PUB Date Free Form | May 71, | |||||||||||
Category Codes | 6L; 58A | |||||||||||
NTIS Prices | PC-GPO/MF A01-NTIS | |||||||||||
Document Type | NT | |||||||||||
Cataloging Source | NTIS/MT | |||||||||||
Control Number | 326521312 | |||||||||||
Origin | NTIS | |||||||||||
Type | CAT |