Main Title |
Response of portable VOC analysers to chemical mixtures : final report / |
Author |
DuBose, D. A. ;
Brown, G. E. ;
Harris, G. E.
|
Other Authors |
|
CORP Author |
Radian Corp., Austin, TX.;Industrial Environmental Research Lab., Research Triangle Park, NC. |
Publisher |
U.S. Environmental Protection Agency, Research and Development, Industrial Environmental Research Laboratory , |
Year Published |
1981 |
Report Number |
EPA 600-2-81-110; EPA-68-02-3171 |
Stock Number |
PB81-234262 |
OCLC Number |
37473385 |
Subjects |
Air--Pollution--Management ;
Organic compounds--Analysis
|
Additional Subjects |
Gas detectors ;
Calibrating ;
Design criteria ;
Performance evaluation ;
Mathematical models ;
Air pollution ;
Portable equipment ;
Volatile organic compounds ;
Air pollution detection ;
Century systems OVA-108 analyzer ;
Bacharach TLV sniffer analyzer
|
Holdings |
Library |
Call Number |
Additional Info |
Location |
Last Modified |
Checkout Status |
ESAD |
EPA 600-2-81-110 |
|
Region 10 Library/Seattle,WA |
08/15/1997 |
NTIS |
PB81-234262 |
Some EPA libraries have a fiche copy filed under the call number shown. |
|
07/26/2022 |
|
Collation |
54 p. : ill. ; 28 cm. |
Abstract |
The report gives the responses of two types of portable VOC analyzers (Century Systems OVA-108 and Bacharach TLV Sniffer), calibrated with methane and used to measure a variety of chemical vapor mixtures. Instrument response data for both binary and ternary mixtures of selected chemicals are presented. Various empirical models were evaluated to determine an appropriate method of estimating mixture concentration based on instrument response. The evaluation concluded that the instrument response for a mixture falls between the responses expected for the pure compounds in the mixture. Thus, an interpolation or weighted average model can be used to predict the response for mixtures based on known responses for individual chemicals. Both linear and logarithmic weighted average models are applied to the data and presented with estimates of accuracy. In general, these models predicted the instrument response to within 30% of the observed value. |
Notes |
Cover title. "EPA Project Officer: Dr. B. A. Tichenor."--Cover. "June 1981." "EPA-600/2-81-110." |
Place Published |
Research Triangle Park, N.C. : |
Supplementary Notes |
See also PB81-234254. |
Corporate Au Added Ent |
Industrial Environmental Research Laboratory (Research Triangle Park, N.C.) |
PUB Date Free Form |
1981. |
NTIS Prices |
PC A04/MF A01 |
BIB Level |
m |
Cataloging Source |
OCLC/T |
OCLC Time Stamp |
19970814181908 |
Language |
ENG |
Origin |
OCLC |
Type |
MERGE |
OCLC Rec Leader |
01193nam 2200289Ka 45010 |