RECORD NUMBER: 755 OF 2473Main Title | Dynamic Dilution System for Auditing Ambient Sulfur Dioxide Analyzers. | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Author | Barnard, William F. ; Brande, Ronald E. ; | |||||||||||
CORP Author | Environmental Monitoring Systems Lab., Research Triangle Park, NC. | |||||||||||
Year Published | 1983 | |||||||||||
Report Number | EPA-600/D-83-115; | |||||||||||
Stock Number | PB83-247452 | |||||||||||
Additional Subjects | Sulfur dioxide ; Monitors ; Auditing ; Gas detectors ; Performance evaluation ; Design criteria ; Concentration(Composition) ; Standards ; Air pollution ; Air pollution detection | |||||||||||
Holdings |
|
|||||||||||
Collation | 21p | |||||||||||
Abstract | This paper discusses the development, evaluation, and field performance of a device designed to provide accurate sulfur dioxide concentration standards suitable for auditing the accuracy of continuous, ambient SO2 monitors. This compact, lightweight, device has been subjected to extensive testing and found to be a reliable and accurate SO2 audit system. A discussion of the evaluation and acceptance testing as well as the results from two years of field audits is presented. |