Main Title |
Comparison of Collection Procedures for the Reinjection Ion Chromatography of Water. |
Author |
Hoover, T. B. ;
Yager, G. D. ;
|
CORP Author |
Environmental Research Lab., Athens, GA. |
Year Published |
1984 |
Report Number |
EPA/600/J-84/171; |
Stock Number |
PB85-151298 |
Additional Subjects |
Trace elements ;
Water analysis ;
Potable water ;
Comparison ;
Chemical analysis ;
Reprints ;
Ion chromatography ;
Water pollution detection ;
Procedures
|
Holdings |
Library |
Call Number |
Additional Info |
Location |
Last Modified |
Checkout Status |
NTIS |
PB85-151298 |
Some EPA libraries have a fiche copy filed under the call number shown. |
|
07/26/2022 |
|
Collation |
5p |
Abstract |
Four collection protocols for the heart-cutting procedure in ion chromatography were compared statistically to determine the best method for use in the analysis of trace ions in water. In the comparison, the protocols were applied in the analysis of 0.1 to 0.4-ppm AS(V) added to a municipal drinking water. Best precision was found when the collection was based on detector response rather than time and when collection included up to half of the next following major peak (SO4(-2)). |