Main Title |
Innovative technology verification report : Xcalibur ElvaX XRF Analyzer / |
CORP Author |
Tetra Tech EM Inc.; Tetra Tech Environmental Management, Inc., Cincinnati, OH.; Environmental Protection Agency, Las Vegas, NV. National Exposure Research Lab. |
Publisher |
U.S. Environmental Protection Agency, Office of Research and Development, National Exposure Research Laboratory, Environmental Sciences Division, Characterization and Monitoring Branch, |
Year Published |
2006 |
Report Number |
EPA 540-R-06-006 |
Stock Number |
PB2006-109028 |
OCLC Number |
181373427 |
Additional Subjects |
Trace elements ;
X-ray fluorescence analyzer ;
Collection locations ;
Field demonstrations ;
Data collection ;
Evaluation designs ;
Reference laboratories ;
Performance evaluation ;
Economic analysis ;
Appendices ;
Tables(Data) ;
Soils ;
Sediments ;
Costs ;
Measurements ;
Xcalibur ElvaX XRF Analyzer ;
Innovative technology verification reports ;
Target elements
|
Internet Access |
|
Holdings |
Library |
Call Number |
Additional Info |
Location |
Last Modified |
Checkout Status |
ELBD ARCHIVE |
EPA 540-R-06-006 |
Received from HQ |
AWBERC Library/Cincinnati,OH |
10/04/2023 |
NTIS |
PB2006-109028 |
Some EPA libraries have a fiche copy filed under the call number shown. |
|
07/26/2022 |
|
Collation |
xiv, 71, [103] pages : color illustrations, charts (some color) ; 28 cm |
Abstract |
The Elvatech, Ltd. ElvaX (ElvaX) x-ray fluorescence (XRF) analyzer distributed in the United States by Xcalibur XRF Services (Xcalibur), was demonstrated under the U.S. Environmental Protection Agency (EPA) Superfund Innovative Technology Evaluation (SITE) Program. The field portion of the demonstration was conducted in January 2005 at the Kennedy Athletic, Recreational and Social Park (KARS) at Kennedy Space Center on Merritt Island, Florida. The demonstration was designed to collect reliable performance and cost data for the ElvaX analyzer and seven other commercially available XRF instruments for measuring trace elements in soil and sediment. The performance and cost data were evaluated to document the relative performance of each XRF instrument. This innovative technology verification report describes the objectives and the results of that evaluation and serves to verify the performance and cost of the ElvaX analyzer. Separate reports have been prepared for the other XRF instruments that were evaluated as part of the demonstration. The objectives of the evaluation included determining each XRF instrument's accuracy, precision, sample throughput, and tendency for matrix effects. To fulfill these objectives, the field demonstration incorporated the analysis of 326 prepared samples of soil and sediment that contained 13 target elements. |
Notes |
"EPA 540-R-06-006." "April 2006." Cover title. "Contract no. 68-C-00-181, task order no. 42." Includes bibliographical references (page 71). |