Record Display for the EPA National Library Catalog

RECORD NUMBER: 18 OF 132

Main Title Automatic Beaker Sample Changer.
Author Lobrin, L. B. ; Boot, R. L. ;
CORP Author Federal Water Pollution Control Administration, Cincinnati, Ohio. Analytical Quality Control Lab.
Year Published 1970
Report Number EPA-16020; 04192,; 16020-01/70
Stock Number PB-207 765
Additional Subjects ( Chemical analysis ; Automation) ; ( Water analysis ; Automation) ; Performance evaluation ; Tests ; pH ; Fluoride ; Laboratory equipment ; Electrochemistry ; Beaker Butler analyzer
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB-207 765 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 8p
Abstract
The Beaker Butler, manufactured by Vision Laboratories, Inc., Oriskany, New York, is a prototype sample changer that presents liquid samples to a variety of electrodes. To evaluate the potential use of this equipment for routine analyses, samples were analyzed using this equipment as well as manual methods. The results, which are included in tables, show that the automated step gave results comparable to manual methods with significant reduction in cost and man hours. (Author)