Record Display for the EPA National Library Catalog

RECORD NUMBER: 25 OF 64

Main Title Evaluation of the Hewlett-Packard 5993B Gas Chromatograph-Mass Spectrometer.
Author Foerst, Denis L. ;
CORP Author Environmental Monitoring and Support Lab.-Cincinnati, OH.
Year Published 1981
Report Number EPA-600/4-81-043;
Stock Number PB81-218315
Additional Subjects Gas chromatography ; Mass spectroscopy ; Water analysis ; Performance evaluation ; Sampling ; Ions ; Water pollution ; Chemical analysis ; Sediments ; Bromine organic compounds ; Computer aided analysis ; Water pollution detection ; Water quality ; EPA method 624 ; EPA method 625 ; NTISEPAORD
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB81-218315 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 53p
Abstract
This report gives the results of an evaluation of a Hewlett-Packard Model 5993B computerized gas chromatograph/mass spectrometer. The evaluation was done according to the procedures found in EPA research report number EPA-600/4-80-025, 'Performance Tests for the Evaluation of Computerized Gas Chromatography/Mass Spectrometry Equipment and Laboratories.' Changes in the 'Performance Test' specifications for the BFB spectrum validation test have been made and are noted in the results for Tests 1 and 8. The system performed satisfactorily during the evaluation and is acceptable for the analyses of organic compound using methods similar to EPA Methods 624 and 625. The HP-5993B system produced valid mass spectra, exhibited good to excellent stability, and gave a satisfactory detection limit. The precision and accuracy for external or internal quantitation using direct sample injection or purge and trap were excellent. Both 14 AMU and PBS mass spectral library searches were accurate; however, the search times using the PBS library search were longer than the requirements 'Performance Tests'. Some problems, particularly with brominated compounds, were discovered during this evaluation. The manufacturer corrected the instrument to alleviate these problems.