Record Display for the EPA National Library Catalog

RECORD NUMBER: 272 OF 297

Main Title Study on Infrared Gas-Filter Correlation Spectrometer for Measuring Low-Concentration Methanol Gases.
Author Cha, S. ; Gabele, P. A. ;
CORP Author Environmental Protection Agency, Research Triangle Park, NC. Atmospheric Sciences Research Lab. ;Northrop Services, Inc., Research Triangle Park, NC. ;Illinois Univ. at Chicago Circle.
Year Published 1986
Report Number EPA/600/J-86/422;
Stock Number PB88-140462
Additional Subjects Gas filters ; Spectrometers ; Performance ; Carbinols ; Exhaust emissions ; Infrared analysis ; Measurement ; Vapors ; Reprints ; Gas filter correlation spectrometer ; Absorption spectra
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB88-140462 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 7p
Abstract
A gas-filter correlation spectrometer that employed a wide spectral band of infrared radiation and gas-filter cell has been built and successfully tested to measure methanol concentrations up to 113 ppm. The instrument demonstrates a detection limit of approximately 1.0 ppm. The interference of nonmethanol gases is negligible when automobile exhausts are used for the test. The most prominent, though not serious, problem appears to be the signal drift, which is believed to be improved by thermally insulating the instrument. The analyzer demonstrated its capability for real-time analysis of emissions from methanol-fueled vehicles due to its simplicity and fast response as compared with wet chemical methods. (Copyright (c) 1986 Society of Photo-Optical Instrumentation Engineers.)