Record Display for the EPA National Library Catalog

RECORD NUMBER: 47 OF 245

Main Title Emanation System for Determining Small Quantities of Radium-226.
Author Blanchar, Richard L. ;
CORP Author Public Health Service, Washington, D.C. Div. of Radiological Health.
Year Published 1964
Report Number PHS-Pub-999-RH-9;
Stock Number PB-189 345
Additional Subjects ( Radium ; Microanalysis) ; ( Radon ; Scintillation counters) ; Chemical analysis ; Radioactive isotopes ; Radium 226
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB-189 345 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 26p
Abstract
A detailed account is given of the construction, operation, and characteristics of an emanation system for the determination of radium-226. The method is specific for radium-226, is well suited for large volume samples containing small quantities of this nuclide, and will accommodate any type of sample that can be prepared in an aqueous solution. The limit of detection at the 95-percent confidence level was calculated to be 0.016 picograms of radium-226, and the transfer and collection of radon-222 was determined to be 100 percent with an average deviation of 3 percent. (Author)