Main Title Comparison of Priority Pollutant Response Factors for Triple and Single Quadrupole Mass Spectrometers.
Author Sauter, A. D. ; Betowski, L. D. ; Ballard, J. M. ;
CORP Author Environmental Monitoring Systems Lab., Las Vegas, NV. ;Lockheed Engineering and Management Services Co., Inc., Las Vegas, NV.
Year Published 1983
Report Number EPA/600/J-83/303;
Stock Number PB85-219368
Additional Subjects Mass spectrometers ; Quality assurance ; Environmental surveys ; Stability ; Chemical analysis ; Gas chromatography ; Performance evaluation ; Comparison ; Laboratory equipment ; Standards ; Reprints ; Relative sensitivity factors ; Quadrupole mass spectrometers
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB85-219368 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 6p
Abstract
Seventy-four percent of the electron impact GC/MS response factors (RF) determined on a triple quadrupole mass spectrometer for 53 extractable priority pollutants were found to be within + or - 15 percent of values determined in an independent interlaboratory single quadrupole GC/MS study. Furthermore, the RF values were shown to be independent of whether quadrupole Q1 or quadrupole Q3 was scanned. The precision of RF determinations for 53 extractable priority pollutants (mean relative standard deviation 11.9 percent) was found to be similar to that previously published for routine GC/MS multianalyte RF determinations.