Record Display for the EPA National Library Catalog

RECORD NUMBER: 346 OF 677

Main Title Inductively Coupled Plasma-Atomic Emission Spectrometric Analysis of Environmental Samples Using Ultrasonic Nebulization.
Author Taylor, Charles E. ; Floyd, Terrance L. ;
CORP Author Environmental Research Lab., Athens, GA.
Year Published 1981
Report Number EPA-600/J-80-044;
Stock Number PB82-118209
Additional Subjects Atomic spectroscopy ; Ultrasonic radiation ; Sampling ; Laboratory equipment ; Chemical analysis ; Water analysis ; Concentration(Composition) ; Reprints ; Plasma spectroscopy ; Ultrasonic nebulizers ; Water pollution detection
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB82-118209 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 8p
Abstract
Ultrasonic and pneumatic nebulizers used in inductively coupled plasma-atomic emission spectrometry (ICP-AES) are compared in the analysis of standard materials and environmental samples. Results verify that ICP-AES with ultrasonic nebulization is capable of lower detection limits than ICP-AES with a pneumatic system. The ultrasonic system used with desolvation, therefore, is the more desirable system when ultratrace ICP-AES quantitative analyses are required in a variety of environmental samples.