Main Title Evaluation of the Westinghouse Capacitor Discharge Mobile X-ray Machine.
Author Grove, J, G. F. ;
CORP Author Bureau of Radiological Health, Rockville, Md. Div. of Electronic Products.
Year Published 1971
Report Number BRH/DEP-71-8;
Stock Number PB-200 400
Additional Subjects ( X-ray apparatus ; Radiation hazards) ; ( Health physics ; X-ray apparatus) ; Dosimetry ; Exposure ; Radiation dosage ; Public health ; Capacitors ;
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB-200 400 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 32p
Abstract
Tests were performed on a Westinghouse model MO-120C capacitor discharge mobile x-ray machine to determine accuracy of voltage indications, the relationship between voltage setting and current-time, line voltage effects and current drain, exposure reproducibility, focal spot size and location, correspondence between the collimator and light localizer, inherent filtration, half-value layer, radiation output, and tube housing and collimator leakage. Spurious radiation was found to originate in the tube under certain conditions. These conditions were investigated, and the observed radiation levels are reported. (BRH/DEP)