Record Display for the EPA National Library Catalog

RECORD NUMBER: 1 OF 1

Main Title Evaluation of the Toshiba Capacitor Discharge Mobile X-Ray Machine.
Author Grove, J, G. F. ;
CORP Author Bureau of Radiological Health, Rockville, Md. Div. of Electronic Products.
Year Published 1971
Report Number BRH/DEP-72-1; 72-8006;
Stock Number PB-202 925
Additional Subjects ( X-ray apparatus ; Evaluation) ; ( Radiation hazards ; X-ray apparatus) ; Safety engineering ; Mobile equipment ; Spurious radiation ; Capacitor discharge X-ray machines
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB-202 925 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 30p
Abstract
Tests were performed on a Toshiba KCD-10M-6 capacitor discharge mobile x-ray machine to determine accuracy of voltage indications, the relationship between voltage setting and current-time values, line voltage effects and current drain, exposure reproducibility, focal spot size and location, correspondence between the collimator and light localizer, inherent filtration, half-value layer, radiation output, and tube housing and collimator leakage. Spurious radiation was found to originate in the tube under certain conditions. These conditions were investigated, and the observed radiation levels are reported. (Author)