Record Display for the EPA National Library Catalog

RECORD NUMBER: 15 OF 50

Main Title Design and Evaluation of an Electronic Radiography System.
Author Miller, Jr, William B. ; Baker, II, Virgil D. ; Ween, Heinz S. ; Laros, James H. ; Colvi, Richard S. ;
CORP Author Bureau of Radiological Health, Rockville, Md. Div. Medical Radiation Exposure.
Year Published 1971
Report Number BRH/DMRE-72-3; PH-86-66-153; 72-8014;
Stock Number PB-204 930
Additional Subjects ( Fluoroscopes ; Television display systems) ; ( Radiation dosage ; Fluoroscopy) ; Radiography ; Video signals ; Magnetic recording ; Magnetic disks ; Image tubes ; Storage tubes ; Magnetic storage ; Digital systems ;
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB-204 930 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 80p
Abstract
The report describes the development and evaluation of a low-exposure electronic radiography system with these major features: (1) reduction in x-ray photon requirements by a factor of approximately 100 in comparison with the usual high speed film-screen combination; (2) instant x-ray imaging and storage for immediate diagnosis, using an exposure time of approximately 1/60 second; and (3) adaptability to the usual televised x-ray image intensifier system. (Author)