Record Display for the EPA National Library Catalog
RECORD NUMBER: 5 OF 28Main Title | Comparison of Gas Chromatography/Fourier Transform Infrared Spectrometric Gram-Schmidt Reconstructions from Different Interferometers. | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Author | Brissey, G. M. ; Henry, D. E. ; Giss, G. N. ; Yang, P. W. ; Griffiths, P. R. ; | |||||||||||
CORP Author | California Univ., Riverside. Dept. of Chemistry.;National Science Foundation, Washington, DC.;Environmental Research Lab., Athens, GA. Office of Research and Development. | |||||||||||
Publisher | c1984 | |||||||||||
Year Published | 1984 | |||||||||||
Report Number | NSF-CHE82-08073; EPA/600/J-84/421; | |||||||||||
Stock Number | PB89-208904 | |||||||||||
Additional Subjects | Gas chromatography ; Interferometers ; Infrared spectroscopy ; Fourier transformations ; Comparison ; Graphs(Charts) ; Reprints ; | |||||||||||
Holdings |
|
|||||||||||
Collation | 7p | |||||||||||
Abstract | The results of an experimental comparison of Gram-Schmidt GC/FT-IR chromatogram reconstructions as calculated from data measured on two different interferometers are presented. The comparison was made using the same sample, GC column, separation conditions, transfer lines, light pipe, detector, and detector optics. Only the interferometers, source, source optics, and data systems were varied. It is determined that the optimum vector displacement is dependent on the stability of the interferometer used, but in general the centerburst should be just outside the sampled area. (Copyright (c) 1984 American Chemical Society.) |