Record Display for the EPA National Library Catalog

RECORD NUMBER: 8 OF 10

Main Title Field-Deployable Monitors for Volatile Organic Compounds in Air.
Author Berkley, R. E. ;
CORP Author Environmental Protection Agency, Research Triangle Park, NC.
Publisher c1994
Year Published 1994
Report Number EPA/600/J-94/353;
Stock Number PB94-210424
Additional Subjects Air pollution monitoring ; Organic compounds ; Portable equipment ; Air pollution sampling ; Gas chromatography ; Performance evaluation ; Quality control ; Field tests ; Reprints ; VOCs(Volatile organic compounds)
Holdings
Library Call Number Additional Info Location Last
Modified
Checkout
Status
NTIS  PB94-210424 Some EPA libraries have a fiche copy filed under the call number shown. 07/26/2022
Collation 12p
Abstract
Volatile organic compounds in ambient air are usually estimated by trapping them from air or collecting whole air samples and returning them to a laboratory for analysis by gas chromatography using selective detection. Data do not appear for several days, during which sample integrity could become compromised. Immediate data can be obtained, and sampling errors minimized, by analyzing with a field-deployable instrument at the time samples are collected. Portable gas chromatographs are available, but they don't fully meet the need for quick, high-quality data under field conditions. Improved waterproofing, temperature regulation, and energy efficiency are particularly crucial to true field-detection, high-speed chromatography, polycapillary chromatography, and peak modulation may lead to useful enhancements in future.