QUALITY ASSURANCE CONSIDERATIONS FOR USE OF THE FLUORIMAGER SI AND FRAGMENT ANALYSIS SOFTWARE

Citation:

Christ, S A., R N. Silbiger, M. Garg, S E. Franson, AND G P. Toth. QUALITY ASSURANCE CONSIDERATIONS FOR USE OF THE FLUORIMAGER SI AND FRAGMENT ANALYSIS SOFTWARE. ELECTROPHORESIS 21(5):874-888, (2000).

Description:

The Fluorimager SI (FSI) from Molecular Dynamics is one of several scanning instruments available for the detection of fluorescent emissions associated with DNA samples in a variety of matrices (agarose and polyacrylamide gels, membranes and microplates). In our laboratory, we measured the electrophoretic moblity of randomly amplified polymorphic DNA (RAPD) fragments stained with ethidium bromide in agarose using the FSI to scan gels and the associated Molecular Dynamics software (ImageQuaNT and FragmeNT Analysis) for analysis. Initial scans and analyses resulted in inconsistent band detection across the same gel and across several scans of the same gel. To determine the best types of calibration for the instrument, several factors were considered and then evaluated. Tests of calibration acceptability were also evaluated. Band detection by FragmeNT Analysis was improved following optimization of matrices and parameters used in calibration and experimental scans. In addition, use of software templates for analysis and modifications in the staining procedure, which have resulted in decreased instrument associated variance, are discussed.