You are here:
PULSATION DAMPER SOLVES ANALYZER PROBLEM (JOURNAL VERSION)
Citation:
Montgomery, J. PULSATION DAMPER SOLVES ANALYZER PROBLEM (JOURNAL VERSION). U.S. Environmental Protection Agency, Washington, D.C., EPA/600/J-88/208 (NTIS PB89134795), 1988.
Description:
The paper describes the use of a pulsation damper to solve a severe output signal interference problem with a Beckman Model 402 Total Hydrocarbon Analyzer. The analyzer was used in a computerized data-gathering system. The problem was traced to compressions/decompressions of flow caused by the sample pump.
Record Details:
Record Type:DOCUMENT( REPORT )
Product Published Date:12/31/1988
Record Last Revised:12/22/2005
Record ID:
50612