Science Inventory

DIFFERENTIAL PULSE POLAROGRAPHY OF PHENYLARSINE OXIDE

Citation:

Lowry, J., R. Smart, AND K. Mancy. DIFFERENTIAL PULSE POLAROGRAPHY OF PHENYLARSINE OXIDE. U.S. Environmental Protection Agency, Washington, D.C., EPA/600/J-78/188.

Description:

Differential pulse polarography was applied to analyze for phenylarsine oxide (PAO) in aqueous solutions at different pH values. Optimization of the instrumental artifacts resulted in a detection limit of 10 to the -8th power M PAO at pH 7.3, a relative standard deviation of 1.7%, and a maximum sensitivity of 450 microampere/mM PAO. The polarographic reduction of phenylarsine oxide was found to be pH dependent. Cyclic voltammetry and coulometry were used to characterize the electrode process.

Record Details:

Record Type:DOCUMENT( REPORT )
Product Published Date:05/24/2002
Record Last Revised:04/16/2004
Record ID: 50301