Office of Research and Development Publications

DETERMINATION OF TRACE ANIONS IN WATER BY MULTIDIMENSIONAL ION CHROMATOGRAPHY

Citation:

Hoover, T. AND G. Yager. DETERMINATION OF TRACE ANIONS IN WATER BY MULTIDIMENSIONAL ION CHROMATOGRAPHY. U.S. Environmental Protection Agency, Washington, D.C., EPA/600/J-84/002 (NTIS PB84175256).

Description:

Selenate, selenite, and arsenate ions were separated from the major anions chloride, nitrate, and sulfate in drinking water, surface water, and groundwater sources by collecting a selected portion of the ion chromatogram, after suppression, on a concentrator column and reinjecting at the original chromatographic conditions. Statistical detection limits varied from 0.02 to 1.2 micrograms of trace element depending on the minor components to be separated and on the water matrix but independent of initial sample size from 2 to 10 ml. The maximum reliably separated molar ratio was 1300 for sulfate/selenate in well water. Carbonate-bicarbonate eluent compositions were optimized for each trace ion.

Record Details:

Record Type:DOCUMENT( REPORT )
Product Published Date:05/24/2002
Record Last Revised:12/10/2002
Record ID: 49980