Science Inventory

A SIMPLE PHOTOMETER FOR PRECISE DETERMINATION OF DISSOLVED OXYGEN CONCENTRATION BY THE WINKLER METHOD WITH RECOMMENDATIONS FOR IMPROVING RESPIRATION RATE MEASUREMENTS IN AQUATIC ORGANISMS

Citation:

Vargo, S. AND R. Force. A SIMPLE PHOTOMETER FOR PRECISE DETERMINATION OF DISSOLVED OXYGEN CONCENTRATION BY THE WINKLER METHOD WITH RECOMMENDATIONS FOR IMPROVING RESPIRATION RATE MEASUREMENTS IN AQUATIC ORGANISMS. U.S. Environmental Protection Agency, Washington, D.C., EPA/600/J-81/531 (NTIS PB82181843), 1981.

Description:

A simple inexpensive photometer designed for Winkler titration end-point detection is described. The precision of replicate dissolved oxygen measurements using this instrument was 0.06-0.22%. This high precision is needed to measure the small changes in dissolved oxygen concentration for determining the respiration rate of small aquatic organisms. Other recommendations for improving precision are also presented.

Record Details:

Record Type:DOCUMENT( REPORT )
Product Published Date:12/31/1981
Record Last Revised:12/22/2005
Record ID: 43255