Science Inventory

EFFECT OF SATELLITE LINES FROM X-RAY SOURCE ON X-RAY DIFFRACTION PEAKS

Citation:

Briden, F. AND D. Natschke. EFFECT OF SATELLITE LINES FROM X-RAY SOURCE ON X-RAY DIFFRACTION PEAKS. U.S. Environmental Protection Agency, Washington, D.C., EPA/600/J-91/196.

Description:

The article discusses the development of a method for relating reactivity to crystallite size and strain parameters obtained by the Warren-Averbach technique. PA has been using crystallite size and strain data obtained from x-ray diffraction (XRD) peak profile analysis to predict the reactivity of solid calcium hydroxide sorbent with acid gases in combustion streams. he software used for the calculations requires that the XRD peak profile be corrected with a distribution function before application in the Warren-Averbach analysis. he curve-fitting software, available on the system, includes the Gaussian, Lorentzian, pseudo-Voight, Pearson VII, and Split Pearson VII distribution functions. he function parameters are specified roughly at the start, and the computer algorithm then optimizes the fit parameters by the least squares method. ost of the work has been on calcium hydroxide sorbent materials with calcium fluoride as a standard to account for instrument effects on the peaks. t has been observed that none of the functions produce a satisfactory fit for the experimental XRD peaks.

Record Details:

Record Type:DOCUMENT( REPORT )
Product Published Date:05/24/2002
Record Last Revised:04/16/2004
Record ID: 41314