Main Title Environmental technology verification report: PCB detection technology : Dexsil Corporation L2000DX analyzer /
Author Dindal, Amy B.; Banyne, C. K.; Jenkins, R. A.; Koglin, E. N.
CORP Author Oak Ridge National Lab., TN.;Environmental Protection Agency, Las Vegas, NV. National Exposure Research Lab.
Publisher U.S. Environmental Protection Agency, Office of Research and Development, National Exposure Research Laboratory, Environmental Sciences Division,
Place Published Las Vegas, NV. :
Year Published 2001
Report Number EPA/600/R-01/049; NERL-LV-01-085
Stock Number PB2002-100440
OCLC Number 48042451
Subjects PCB(Polychlorinated biphyenyls); Polychlorinated biphyenyls; Detectors; Chlorinated solvents; Transformers; Electrodes; Technology assessment; Soil pollution; Portable equipment
Internet Access
Description Access URL
http://nepis.epa.gov/Exe/ZyPURL.cgi?Dockey=P1001NNO.txt
Holdings
Library   Call Number Additional Info Location Date Modified
ERBD EPA/600/R-01/049 NERL/ESD-LV Library/Las Vegas,NV 10/05/2001
NTIS PB2002-100440 Most EPA libraries have a fiche copy filed under the call number shown. Check with individual libraries about paper copy. NTIS 01/01/1988
Collation 30 p. ; ills. : 28 cm.
Abstract The L2000DX Analyzer (dimensions, 9 x 9.5 x 4.25 in.) is a field-portable ion-specific electrode instrument, weighing approximately 5 lb 12 oz, designed to quantify concentrations of PCBS, chlorinated solvents, and pesticides in soils, water, transformer oils, and surface wipes. The L2000DX can be operated in the field powered by a rechargeable 8-V gell cell, or in the laboratory using 120-V AC power. To prepare a sample for analysis, 5 mL of the oil is collected in a polyethylene reaction tube. Two glass ampules contained in the reaction tube are broken to introduce metallic sodium to the oil. The mixture is then shaken for 10 s and allowed to react for a total of 1 min. The sodium strips the covalently bonded chlorine atoms off the PCB molecule. An aqueous extraction solution is added to the reaction tube to adjust the pH, destroy the excess sodium and extract and isolate the newly formed chloride ions in a buffered aqueous solution. The aqueous layer is decanted, filtered, and collected in an analysis vial. The ion-specific electrode is put into this aqueous solution to measure the millivolt potential. The potential is then converted to the equivalent PCB concentration. The lowest concentration reported by the L2000DX is typically 3 ppm. The performance of a previous version of this instrument (the L2000 PCB/Chloride Analyzer) was verified by ETV for soil and solvent extracts in 1998.
Notes Cover title. "August 2001." "EPA/600/R-01/049."
Supplementary Notes Sponsored by Environmental Protection Agency, Las Vegas, NV. National Exposure Research Lab.
Availability Notes Product reproduced from digital image. Order this product from NTIS by: phone at 1-800-553-NTIS (U.S. customers); (703)605-6000 (other countries); fax at (703)605-6900; and email at orders@ntis.gov. NTIS is located at 5285 Port Royal Road, Springfield, VA, 22161, USA.
Author Added Ent
Bayne, Charles K.;
Jenkins, Roger A.;
Koglin, Eric N.
Corporate Au Added Ent National Exposure Research Laboratory Environmental Sciences Division, Las Vegas, Nev. ; United States. Environmental Protection Agency. Office of Research and Development. ; Oak Ridge National Laboratory.
Alternate Title PCB detection technology. Dexsil Corporation L2000DX analyzer.
PUB Date Free Form 2001.
NTIS Prices PC A04/MF A01
BIB Level m
OCLC Time Stamp 20010927145216
Cataloging Source OCLC/T
Language eng
Origin OCLC
Type MERGE
OCLC Rec Leader 01301nam 2200313Ka 45020